The Label Count That Does Not Close: FDA Labeling Controls on Top of ISO 13485

Under QMSR, 21 CFR 820.45 still adds labeling and packaging controls to ISO 13485. Where kit label mixups happen, what reconciliation shows, and what to ask a kit manufacturer.
Supera Fulfillment card reading Label Controls with a line drawing of a stacked label and a check mark

5 min read

In this guide
  1. What 820.45 adds to ISO 13485
  2. The five items, and where each goes wrong on a kit
  3. Mixups happen at changeover
  4. What reconciliation looks like
  5. Inspection is not the same as sampling
  6. Questions to ask a kit manufacturer

FDA’s Quality Management System Regulation took effect February 2, 2026 and pulled ISO 13485 into U.S. device law by reference. Most of the old Part 820 went away. Labeling did not. Section 820.45 still sits on top of ISO 13485, and it asks for something the standard alone does not spell out: proof that the right label went on the right kit, checked before use, with the result written down.

For at-home testing and specimen collection kits, that matters more than it does for a single device. A kit carries several labels at once: the outer carton, the inner tray or bag, the collection device, the return mailer, and the instructions. Each one can carry a lot number, an expiration date or a device identifier. Each one is a chance for a mixup.

What 820.45 adds to ISO 13485

ISO 13485 clause 7.5.1 covers control of production in general terms. Section 820.45 opens with “in addition to the requirements of Clause 7.5.1” and then gets specific. It asks for a documented procedure that keeps labeling and packaging integrity, inspection, storage and operations under control through processing, storage, handling and distribution. It then sets three requirements.

  1. Accuracy check before release or storage. Labeling and packaging are examined for accuracy, covering at least five named items.
  2. Documented release. Release of labeling for use is documented under ISO 13485 clause 4.2.5.
  3. Mixup prevention. Labeling and packaging operations are set up and maintained to prevent mixups, including inspection before use to confirm the labeling matches the medical device file, with the results documented.

The QMSR overview covers the wider change. This post stays on the label, because it is where a quality system that looks complete on paper most often breaks on the floor.

The five items, and where each goes wrong on a kit

820.45(a) itemHow it fails on a multi-component kit
UDI, UPC or other device identificationA carton printed for one kit configuration is applied to a near-identical configuration with a different component.
Expiration dateThe kit date is set from the box stock instead of the shortest-dated component.
Storage instructionsThe carton says one range while a component inside carries a narrower one.
Handling instructionsAn old instruction insert revision stays at the station after a change.
Additional processing instructionsA return or shipping instruction does not match the return path the program now uses.
Examples of failure modes, not findings from any specific program.

None of these is a printing defect. The printer did its job. Each is a correct label in the wrong place, which is exactly what an artwork approval cannot catch. Artwork control, covered in the dieline and artwork files guide, proves the design is right. Section 820.45 asks you to prove the application was right.

Mixups happen at changeover

Labels rarely go wrong in the middle of a steady run. They go wrong when the line switches from one lot, one configuration or one client to the next, and something from the last job stays behind. A partial roll on the printer. A stack of inserts under the bench. A carton pallet that was not pulled back to the cage.

A label count that does not close is a mixup you have not found yet.

The controls that address this are plain, and most of them are counts and signatures:

  • Line clearance. Before a run starts, someone confirms the area holds no labeling, components or finished kits from the prior job, and signs for it. A second person verifies.
  • Controlled issuance. Labels are issued to the line by quantity against the batch record, not pulled from an open shelf.
  • First-piece check. The first assembled kit is checked against the approved label set and the batch record before the run continues.
  • Reconciliation. At the end of the run, issued labels are counted against applied, destroyed and returned. A gap gets investigated before release.
  • Segregation. Look-alike labels for different configurations are never staged at the same station at the same time.

What reconciliation looks like

Carton labels, one lot (illustrative)Count
Issued to the line5,200
Applied to released kits5,000
Destroyed (misprints, setup)150
Returned to controlled storage40
Unaccounted10
Illustrative example.

Ten labels out of 5,200 looks like noise. It is not. Ten labels can be on ten kits from another lot, sitting in a warehouse with the wrong expiration date or the wrong device identifier. Until the count closes or the gap is explained, the lot should not ship. The investigation is usually short: a miscounted destroy bag, a roll end that went back to storage uncounted. The point is that someone looks, and the batch record shows they looked.

Reconciliation also pays off later. If a component is ever pulled, the records that let you trace a recall back through kit lots are only as good as the proof that each lot carried its own labels.

Inspection is not the same as sampling

Section 820.45 asks for inspection of labeling before use and documentation of the result. It does not turn every kit into a full inspection. Finished-kit acceptance still runs on the sampling plan the procedure sets, as covered in AQL and sampling plans. The label controls are what make sampling defensible: if the right labels were the only labels at the station, a sample can tell you something about the lot.

Questions to ask a kit manufacturer

  • Show me a completed line clearance record from a recent changeover. Who signed, and who verified?
  • How are labels issued to the line, and what record ties the issued quantity to the batch?
  • What happens when the label reconciliation does not close? Show me one where it did not.
  • Where in the batch record is the first-piece label check, and what is it checked against?
  • How do you keep look-alike labels for two configurations apart?
  • Who releases labeling for use, and where is that release documented?

Whether 820.45 applies to your kit, and to whom, depends on the product’s regulatory status and each party’s role. That is a regulatory determination. Confirm it with your regulatory lead or counsel rather than relying on a general guide.

At Supera Fulfillment, kits are built under our ISO 13485 certified quality system in an ISO Class 7 cleanroom at our FDA-registered facility. If you are evaluating a kit manufacturer, the questions above are a fair place to start, whoever you work with.

Sources: 21 CFR 820.45, Device labeling and packaging controls (eCFR); FDA, Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496 (Feb. 2, 2024), effective Feb. 2, 2026; ISO 13485:2016, Medical devices, Quality management systems, clauses 4.2.5 and 7.5.1.

Written by

Michael Brown

Michael Brown is Co-Founder and Chief Commercial Officer of Supera Fulfillment, an ISO 13485 certified contract manufacturer and kitting operation in Houston. He scopes and prices specimen collection kit programs, and works mostly on the parts buyers find out about late: bills of materials, regulatory labeling, return paths, and what a device choice does to a kit. He writes these guides to be useful whether or not you ever work with Supera.